116
QC5200 Series User’s Guide
Adding the profile feature to the Feature template
When the profile measurement is complete, press the OK button at the bottom of the Measure Profile
window to create the profile feature in the Features template and close the Measure Profile window.
Measurement results are
shown in the Features
template, Part View win-
dow and Results window.
The Results window
shows the feature iden-
tification number, the
number of points used in
the profile measurement,
the coordinate center of
the feature, the angular
orientation of the feature
and the overall, least ma-
terial and most material
form errors.
The profile feature is
shown in the Feature
Stamp window with
form error whiskers that passed tolerance tests in
the range of 0 to 50% shown in green, 51 to 100%
in yellow. Form errors that fail the tolerance tests
are shown in red.