Operating parameters for measured value output – HEIDENHAIN PT 855 for Turning User Manual
Page 76

II - 5
Measured Value Output
78
Technical Information
POSITIP 855
Fig. 44: Time diagram for measured value out-
put with Ctrl B
Starting measured value output
C t r l B
C t r l B
t
1
t
3
t
2
t
D
R X D
T X D
Starting measured value output with Ctrl B
t
1
: Delay between Ctrl B command and internal latch
t
1
≤
0.5 ms
t
2
: Delay between internal latch and measured value output
t
2
≤
30 ms
t
3
: Time between end of data output and next latch with Ctrl B
t
3
≥
0 ms
t
D
: Duration of measured value output
The duration of measured value output (t
D
) depends on
•
The selected baud rate (BR),
•
The number of axes (M)
•
The number of blank lines (L)
t
D
=
[s]
Operating parameters for measured value output
The following operating parameters will influence measured
value output — regardless of how you start it.
Number of blank lines after each measured value: P51
Number of blank lines after measured value: P51 = 0 to 99
You can also use the signal for measured value output to
influence position display.
Screen display during measured value output: P23
The display is not stopped during measured
value output (Off):
P23 = 0
The display is stopped during measured value output
and remains stopped as long as the switching input
“output measured value” is active (Concrnt):
P23 = 1
The display is stopped but is updated by every
measured value output (Frozen):
P23 = 2
BR
187
×
M + L
×
11