NLEC by importing an NLEC data .txt file
An NLEC data file created by measuring a certified calibration grid with
an ND 12x or other systems capable of generating NLEC data can be
used to provide NLEC correction data instead of measuring a
calibration grid with the target DRO system. The NLEC data file is
loaded from a computer using Hyperterminal or a similar serial
communications application. Hyperterminal is used in this manual.
To import NLEC data to a computer:
Establish communication between the DRO and the computer. See
"Connecting a computer" on page 64.
Open and connect to Hyperterminal. See "Connecting to
Hyperterminal" on page 65.
On the DRO, press MENU>Setup to display the setup menu.
Use the Arrow keys to highlight the NLEC menu item.
Highlight the NLEC choice field.
Press the Data soft key then the Receive soft key. A pop-up window
notifies to “Send NLEC data now...”
In Hyperterminal, click Transfer>Send Text File...
In the Send Text File window, select the NLEC data file to send to
the DRO and click Open.
A pop-up will notify “The NLEC data was successfully loaded.
Reboot the system.” Reboot the DRO by powering-off the unit then
powering-on.
Highlight the NLEC choice field and
press the Data soft key
Press the Receive soft key