Highlight the Standard field and enter the Standard value at the end
of segment 3. In the example, the value is 3.00000. Then highlight
the Observed field and press the Teach soft key. The system will
enter the value measured at the end of the segment. In the example
the Observed value at Station 3 is 2.98000.

Position the crosshairs over the end of segment 4 on the standard,
highlight the Station field again and press the Inc soft key to
advance the station number to 4. This is the last station in the
example.

Highlight the Standard field and enter the Standard value at the end
of segment 4. In the example, the value is 4.00000. Then highlight
the Observed field and press the Teach soft key. The system will
enter the value measured at the end of the segment. In the example
the Observed value at Station 4 is 4.0000.

Repeat the SLEC process for other axes if necessary.

When SLEC data are entered for all the desired axes, highlight the
Enabled field for each axis and press the On soft key to activate SLEC
for each axis.

Press the FINISH key to save parameters and return to the Setup
menu.

Enter Standard value and Teach the
Observed value for Station 1

Enter Standard value and Teach the
Observed value for Station 2

Enter Standard value and Teach the
Observed value for Station 3

Enter Standard value and Teach the
Observed value for Station 4

Press the On soft key when all desired
axes have been compensated