Stage squareness calibration
The Squareness screen contains data and choice fields for calibrating
the squareness of the measuring system. The calibration of stage
squareness requires the use of a certified square artifact.
To calibrate squareness:
Align the squareness calibration artifact with the reference axis for
the squareness calibration. This axis will be the Master axis.
Press MENU>Setup to display the setup menu and then highlight the
Squareness menu item.
Highlight the Master Axis choice field and press a soft key to select
the reference (Master) axis for squareness calibration.
Highlight the Angle data field and press the Teach soft key to begin
the squareness calibration.
Follow instructions displayed on the LCD screen to complete the
calibration.
Press the FINISH key.
Highlight the Squareness menu item
Select a Master Axis
Highlight the Angle data field and press
the Teach soft key