9 error indications, Scale errors, 9 er ro r indications 1.9 error indications – HEIDENHAIN ND 100 User Manual User Manual
Page 55

ND 120 QUADRA-CHEK
55
1
.9 Er
ro
r Indications
1.9 Error Indications
Scale errors
Only input encoder scale errors are reported. Scale errors are indicated
by the presence of graphic bars across the DRO screen instead of
numbers. Scale errors can be caused by a few conditions shown in
this table:
Possible cause
Corrective action
Damaged encoder
reader head
Repair or replace the encoder.
Misaligned
encoder reader
head
Recalibrate the reader head.
Electrical noise on
the encoder input
Verify that the earth ground provided by the
power source is in good condition and is
connected to the power system ground.
Verify that the encoder cable is shielded
and that the shield is connected to ground
at the DRO end.
Verify that the reader head is not generating
electrical noise.
Stage (encoder)
velocity too high
Restrict stage velocity. If the velocity
required for error-free operation is slow,
recalibrate the reader head.
Faulty encoder
input connection
Repair or replace the encoder cable or seek
the assistance of your HEIDENHAIN
distributor.