beautypg.com

Guralp Systems CMG-CD24 User Manual

Page 86

background image

CMG-CD24

Instrument description

Response type

Unit

CMG-3T (120s–50Hz)

CMG­3_120S_50HZ

Vel

CMG-3T (360s–50Hz)

CMG­3_360S_50HZ

Vel

CMG-3T (120s–100Hz)

CMG­3_120S_100HZ

Vel

CMG-3TB (30s–50Hz)

CMG­3B_30S_50HZ

Vel

CMG-3TB (100s–50Hz)

CMG­3B_100S_50HZ

Vel

CMG-3TB (120s–50Hz)

CMG­3B_120S_50HZ

Vel

CMG-3TB (360s–50Hz)

CMG­3B_360S_50HZ

Vel

CMG-3TB (360s–100Hz)

CMG­3B_360S_100HZ

Vel

CMG-3V (30s–100Hz)

CMG­3V_30S_100HZ

Vel

CMG-40T (1s–100Hz)

CMG­40_1S_100HZ

Vel

CMG-40T (2s–100Hz)

CMG­40_2S_100HZ

Vel

CMG-40T (10s–100Hz)

CMG­40_10S_100HZ

Vel

CMG-40T (20s–50Hz)

CMG­40_20S_50HZ

Vel

CMG-40T (30s–50Hz)

CMG­40_30S_50HZ

Vel

CMG-40T (60s–50Hz)

CMG­40_60S_50HZ

Vel

CMG-40T (100s–50Hz)

CMG­40_100S_50HZ

Vel

CMG-5T (DC–100Hz)

CMG­5_100HZ

Acc

CMG-6T (1s–100Hz)

CMG­6_1S_100HZ

Vel

CMG-6T (2s–100Hz)

CMG­6_2S_100HZ

Vel

CMG-6T (10s–100Hz)

CMG­6_10S_100HZ

Vel

CMG-6T (30s–100Hz)

CMG­6_30S_100HZ

Vel

7.1.3 Example files

The calibration information for a CMG-3T weak-motion velocity
sensor might look like the following:

[GURALP-DEMO]

Serial-Nos=T3X99

VPC=3.153,3.147,3.159

G=1010,1007,1002

COILCONST=0.02575,0.01778,0.01774

CALVPC=3.161

CALRES=51000

TYPE=CMG-3T

RESPONSE=CMG-3_30S_50HZ Vel

GRAVITY=9.80122

CMG-5T accelerometers use 1Ω calibration resistors, and their coil

constant is set to unity. For example:

[GURALP-CMG5]

Serial-Nos=T5585

86

Issue C