beautypg.com

3 gps and timing systems – Guralp Systems CMG-CD24 User Manual

Page 56

background image

CMG-CD24

This command is in the extended dictionary; to use it, first issue the

command ok-1 and finish with [seal].

5.2.5 TEMP?

Syntax: TEMP?

Display the current temperature measurement from the internal

thermometer.

5.2.6 ETHER

Syntax: ETHER ENABLE | ETHER DISABLE

Enables or disables the optional Ethernet and Wi-Fi devices on the

CD24.

When the Ethernet device is enabled, data produced by the CD24 will
be sent to the device for transmission across the network, unless you

have plugged a serial cable into the Data Out port of the breakout box.
In this case, data will be sent over the standard RS232 interface only.

This is the default behaviour.

When the Ethernet device is disabled, data will always be sent out
over the standard RS232 interface, and the internal Ethernet/Wi-Fi

module will not be used.

5.3 GPS and timing systems

5.3.1 GPS-TYPE

Syntax: type GPS-TYPE

Tells the CD24 which kind of GPS is attached to it.

type can be one of

0, if no GPS is available, or

2, for attached GPS equipment using the NMEA protocol.

5.3.2 HR-CYCLE

Syntax: interval HR-CYCLE

Sets the interval between GPS fixes. Under normal operation, the

system will power on the GPS system every interval hours and

56

Issue C