Guralp Systems CMG-CD24 User Manual

Page 51

background image

Operator's Guide

10. The script will ask you to fill in sensor calibration parameters

for each component you have selected.

Most data can be found on the calibration sheet for your sensor.
Under Instrument response, you should fill in the sensor

response code for your sensor, according to the table below.
Instrument Type should be set to the model number of the

sensor.

If the file calvals.txt exists in the same directory as
Scream!'s executable (scream.exe), Scream! will look there for

suitable calibration values.

See the Scream! manual for full details of this file.
Alternatively, you can edit the sample calvals.txt file

supplied with Scream!.

11. Click OK. The script will return with a graph showing the

responsivity of the sensor in terms of amplitude and phase plots

for each component (if appropriate.)

March 2011