From the main menu:

1. Each channel is configured individually. To configure either,

enter either 1 (selecting Channel 1) or 2 (selecting Channel 2)

and then press the Enter key.

2. Set the Baud Rate to 19200. This is the default baud rate for the

CD24's digital output. If you change the baud rate in Scream! or
using the terminal, you must change the Baud Rate setting here
to match.

3. The remaining settings can be left at their default values, which

is done by pressing Enter at each prompt until you are returned
to the main menu.

4. When you have finished setting up the module, apply the new

settings by selecting option 9 Save and Exit. The module will

re-boot with the new settings in effect.

7.4.3 Configuring the Ethernet port

To configure the Ethernet port

1. If DHCP is not being used to assign IP addresses, enter the

required address manually. The IP address must be set to a
unique value in the network. Enter each octet and press Enter

to move to the next octet.. The current value is displayed in

March 2011