beautypg.com

2 response codes and units – Guralp Systems CMG-CD24 User Manual

Page 85

background image

Operator's Guide

Field

Value

components, separated by commas. For velocity
sensors, the gain, or sensitivity, is given on the

sensor calibration sheet in V/ms

-1

. The gain of

an accelerometer is expressed on its calibration

sheet in V/ms

-2

. Because the outputs are

differential, these are written as “2 × single-

ended-sensitivity”. Most set-ups (including all
those using DM24mk3 and CD24 digitisers) will

use the doubled value, but older integrated
instruments combining a DM24mk2 with a

sensor will use only the single ended sensitivity
value.

COILCONST

The coil constant for the Z, N/S and E/W sensor
components, in A/ms

-2

, separated by commas.

These are given on the sensor calibration sheet.
For a CMG-5T accelerometer, the values are all

set to unity; use COILCONST=1,1,1 for these.

CALRES

The value of the calibration resistor, in Ω, as

given on the sensor calibration sheet. For a
CMG-5T, use CALRES=1.

CALVPC

The sensitivity of the digitiser's calibration
channel, in V per count, as given on the

μ

digitiser calibration sheet. Older CMG-5TD
instruments do not have calibration input

facilities, and thus CALVPC can be omitted.

GRAVITY

The local acceleration due to gravity at the

installation site, in ms

-2

. This field is used by

Scream! to display streams in physical units. By

default, a standard average g value of 9.80665
ms

-2

is used, but this field can be changed for

greater accuracy if desired.

TYPE 
[optional]

The model number of the instrument, used for

display purposes.

RESPONSE

A string describing the theoretical response of

the instrument, in the form response-type unit.
See the table below for the correct values.

7.1.2 Response codes and units

Instrument description

Response type

Unit

CMG-3T (30s–50Hz)

CMG­3_30S_50HZ

Vel

CMG-3T (60s–50Hz)

CMG­3_60S_50HZ

Vel

CMG-3T (100s–50Hz)

CMG­3_100S_50HZ

Vel

March 2011

85