beautypg.com

Guralp Systems CMG-CD24 User Manual

Page 66

background image

CMG-CD24

1, for a 100 s high pass filter;

2, for a 300 s filter;

3, for a 1000 s filter; or

0, to disable the high pass filter.

Note that this filter will affect the data being output as continuous

streams.

5.5.10 PRE-TRIG

Syntax: time PRE-TRIG

Sets the pre-trigger recording time. time is the number of seconds of

data to output from before a trigger is declared.

5.5.11 POST-TRIG

Syntax: time POST-TRIG

Sets the post-trigger recording time. time is the number of seconds of

data to output after a trigger condition lapses. If an event persists for
some time, all triggering components must fall below the threshold

before the trigger condition will lapse; only then will the post-trigger
period begin.

5.5.12 TRIGGERIN

Syntax: TRIGGERIN ENABLE | TRIGGERIN DISABLE

Enables or disables external trigger input, in digitisers equipped with
this option.

Enabling external trigger input allows you to trigger the CD24 from an

external logic level supplied through its digital output port. This
voltage can be between 5 and 10 V supplied between the Trigger In pin

and signal ground. If the CD24 is triggered externally, it will behave
exactly as if it had generated the trigger itself.

5.5.13 TRIGGEROUT

Syntax: TRIGGEROUT ENABLE | TRIGGEROUT DISABLE

Enables or disables external trigger output, in digitisers equipped with
this option.

66

Issue C