Long term test, Setup and testing – Fluke Biomedical SigmaPace 1000 User Manual
Page 97

Transvenous Pacemaker Testing
Setup
and
Testing
4
4-11
Standard test leads or short jumpers can be used to make connections on other
brands and models of pacemakers.
Long Term Test
Tests are performed for only one channel at a time. The overall stability of the
pacemaker’s pulse rate and amplitude are monitored continuously over an
extended time period. The user can independently input desired test limits for
amplitude and rate (2-20) or use the factory default 10 % limits. Upon
completion of the desired long-term test duration, the user can review
pacemaker output pulses falling outside these test limits.
Setup and Testing
The following describes how to set up and test a transvenous pacemaker.
F1
F2
F3
F4
eyr001.eps
1. From the
Select external pacemaker type
screen, press
F2 INV
to
begin Invasive (transvenous) pacemaker testing.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)