Test options – Fluke Biomedical SigmaPace 1000 User Manual
Page 57

Transcutaneous Pacemaker Testing
Test
Options
3
3-3
Test Options
The scheme for transcutaneous pacemaker testing is shown in Figure 3-1. The
specified (or default) test load is pre-selected with each algorithm. If desired,
you can easily select any test load (50
Ω
to 1550
Ω
in 50
Ω
steps) with each
measurement algorithm.
Ensure that you select a test load within the device manufacturer’s specified
operational range. Device-specific measurement algorithms are selectable for
the following list of manufacturers:
•
Physio-Control Lifepak Series (50
Ω
)
•
HP / Agilent Codemaster (400
Ω
)
•
GE / Marquette Responder Series (300
Ω
)
•
Medical Data Electronics (600
Ω
)
•
Medical Research Laboratories (500
Ω
)
•
Zoll Medical PD-Series (1000
Ω
)
•
Zoll Medical M-Series (100
Ω
)
Tests requiring ECG simulation have three available active outputs:
•
Low Level (2): 5 Lead: RL / N, RA / R, LA / L, LL / F, and V1 / C1
Transthoracic: Across the selected ventricular test load
•
High Level (1): Output Amplitude tracks low level Lead I (1000:1)
+ / - 5.0V-peak maximum amplitude
Connector Type: Subminiature Phone Jack