Transcutaneous pacemaker tests, Demand mode test – Fluke Biomedical SigmaPace 1000 User Manual
Page 27

Introduction and Specifications
Instrument
Specifications
1
1-13
Transcutaneous Pacemaker Tests
Output Pulse Measurement
Test Type ............................................................... Quantitative
Measurement Ranges
Output ................................................................. 4.0 mA to 9.99 mA
10.0 mA to 99.9 mA
100 mA to 250 mA
Accuracy......................................................... ±2 % of reading or ±50 µA (whichever
is greater)
Rate .................................................................... 5.0 PPM to 99.9 PPM
100 PPM to 300 PPM
Accuracy......................................................... ±0.5 % of reading or 0.3 PPM
(whichever is greater)
Width................................................................... 1.00 mS to 9.99 mS
10.0 mS to 99.9 mS
Accuracy......................................................... ±0.5 % of reading or 14 µS (whichever
is greater)
Energy ................................................................ 1µJ to 999 µJ
1 mJ to 999 mJ
1.00 J to 1.99 J
Accuracy......................................................... ±5 % of reading / computation
Demand Mode Test
Test Type ............................................................... Qualitative
Physiological Simulation
Selection ............................................................. Normal Sinus Rhythm (NSR)
generates the complete P-QRS-T
complex.
Amplitude............................................................ (V
peak
): 1.0 mV Lead I.
Mode of Operation
Underdrive .......................................................... NSR @ 85 % of measured pulse
interval / rate
Overdrive ............................................................ NSR @ 115 % of measured pulse
interval / rate
Auxiliary Control................................................... The under / overdrive NSR simulations
can also be independently adjusted in
1 BPM increments.