Fluke Biomedical SigmaPace 1000 User Manual
Page 101

Transvenous Pacemaker Testing
Setup
and
Testing
4
4-15
F1
F2
F3
F4
eyr038.eps
6.
Press
F2 RATE
to toggle between
NORM
and
<20.
Note
The
RATE
<20 is used to measure pacer activity under the rate of 20
PPM. The sampling time is extended to accommodate this lower rate.
7. Press
F4 LOAD
UP
or
DOWN
arrow key to select the desired load for
the selected channel.
8. Press
F1
HOLD
to advance the screen:
F1
F2
F3
F4
eyr016.eps
9. Press
F4
XMIT DATA
to transmit data through the serial port.
10. Press
F1
RELEASE
to release the hold and return to the
HOLD
screen.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)