Continuous mode and dual-channel (av) ecg signal – Fluke Biomedical SigmaPace 1000 User Manual
Page 104

SigmaPace 1000
Operators Manual
4-18
Continuous Mode and Dual-channel (AV) ECG Signal
The following test measures the continuous (or non-demand) pacemaker’s
ability to interact with a simulated dual-channel (AV) ECG signal.
F1
F2
F3
F4
eyr032.eps
1. From the
SELECT TEST
menu, press
F3
ASYNC MODE
.
F1
F2
F3
F4
eyr068.eps
2. Press
F4 LOAD
UP
or
DOWN
arrow key to select the load.
3. Press
F1
START
to advance the screen and start the test.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)