beautypg.com

Av interval (delay time), Demand mode test – Fluke Biomedical SigmaPace 1000 User Manual

Page 32

background image

SigmaPace 1000

Operators Manual

1-18

Voltage ................................................................ 0.050 V

peak

to .999 V

peak

1.00 V

peak

to 9.99 V

peak

10.0 V

peak

to 30.0 V

peak

Accuracy..........................................................2 % of reading or 0.05 V

peak

(whichever is greater)

Polarity Indicator..................................................+ or -
Energy .................................................................1 nJ to 999 nJ

1 µJ to 999 µJ – add “J” range

Accuracy..........................................................5 % of reading / computation

Displayed Data

Single: Amplitude - Current

Amplitude ........................................................Volts
Rate
Width
Energy

Dual: Atrial and Ventricular

Amplitude ........................................................Current
Rate
Width
AV Interval

AV Interval (Delay Time)

Test Type................................................................Quantitative
Measurement Ranges

10.0 mS to 99.9 mS
100 mS to 999 mS
Start / Stop Points

Start.................................................................Leading edge of atrial pacemaker

pulse

Stop .................................................................Leading edge of ventricular

pacemaker pulse

Accuracy..............................................................1 % of reading / computation

Demand Mode Test

Test Type................................................................Qualitative
Channels

Single ..................................................................Atrial or Ventricular Only
Dual .....................................................................A+V