Av interval (delay time), Demand mode test – Fluke Biomedical SigmaPace 1000 User Manual
Page 32
SigmaPace 1000
Operators Manual
1-18
Voltage ................................................................ 0.050 V
peak
to .999 V
peak
1.00 V
peak
to 9.99 V
peak
10.0 V
peak
to 30.0 V
peak
Accuracy..........................................................2 % of reading or 0.05 V
peak
(whichever is greater)
Polarity Indicator..................................................+ or -
Energy .................................................................1 nJ to 999 nJ
1 µJ to 999 µJ – add “J” range
Accuracy..........................................................5 % of reading / computation
Displayed Data
Single: Amplitude - Current
Amplitude ........................................................Volts
Rate
Width
Energy
Dual: Atrial and Ventricular
Amplitude ........................................................Current
Rate
Width
AV Interval
AV Interval (Delay Time)
Test Type................................................................Quantitative
Measurement Ranges
10.0 mS to 99.9 mS
100 mS to 999 mS
Start / Stop Points
Start.................................................................Leading edge of atrial pacemaker
pulse
Stop .................................................................Leading edge of ventricular
pacemaker pulse
Accuracy..............................................................1 % of reading / computation
Demand Mode Test
Test Type................................................................Qualitative
Channels
Single ..................................................................Atrial or Ventricular Only
Dual .....................................................................A+V