Refractory period test (ventricular channel), Sigmapace 1000 – Fluke Biomedical SigmaPace 1000 User Manual
Page 38

SigmaPace 1000
Operators Manual
1-24
Physiological Simulation
Selection (Default)...............................................Single Channel Simulation (A)
Square Wave
Atrial Channel......................................................Simulated P-wave
Width ...................................................................1.0 mS
Amplitude ............................................................20.0 mV
peak
Active Outputs .....................................................Atrial Channel (4 mm Banana Jacks)
Only
Additional Waveform Selections
Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO) (Fixed Width 2 mS rise time / 13 mS fall time)
Amplitude
Range...................................................................05 mV
peak
(50 V
peak
) to 50.0 mV
peak
Accuracy..............................................................±5 % of setting
Resolution (Step Size).........................................0.05 mV
peak
steps: 0.05 mV
peak
to 0.95
mV
peak
0.50 mV
peak
steps: 1.0 mV
peak
to 49.5
mV
peak
Width
Range..................................................................0.15 mS to 95.0 mS
Accuracy..............................................................±5 % of setting
Selection Count ...................................................50
Resolution (Step Size).........................................0.05 mS steps: 0.15 mS to 0.95 mS
1.0 mS steps: 1.0 mS to 19.0 mS
5.0 mS steps: 20 mS to 95.0 mS
Active Output.......................................................Atrial Channel (4 mm Banana Jacks)
Only
Intended Pacemaker Types ................................AAI (Atrial Pace and Sense Only)
Compatible Pacemaker Rates.............................30 to 200 PPM
Available Test Load.............................................500
Ω
±1 %
Refractory Period Test (Ventricular Channel)
Test Selections
Paced Refractory Period (PRP)
Sensed Refractory Period (SRP)
Period .....................................................................20 to 500 mS
Accuracy ................................................................±5 % of reading or 1 mS whichever is
greater