Transvenous pacemaker tests – Fluke Biomedical SigmaPace 1000 User Manual
Page 31

Introduction and Specifications
Instrument
Specifications
1
1-17
Pulse Width......................................................... 40 mS
Rate .................................................................... Double pulse: interactive with applied
pacemaker pulse activity in the
demand mode of operation.
Amplitude .............................................................. 1.0 m V
peak
Lead I
Active Outputs ...................................................... Five Lead ECG and selected
ventricular channel test load. This
waveform is present on both outputs.
Pacemaker Compatibility
Compatible Pulse Rates ..................................... 30 to 200 BPM
Intended Pacemaker Type.................................. VVI (Ventricular Pace and Sense
Only)
Transvenous Pacemaker Tests
Output Pulse Measurement
Test Type ............................................................... Quantitative
Display Formats
Single.................................................................. Atrial or Ventricular only
Dual .................................................................... Atrial and ventricular Channels
Measurement Ranges
Current................................................................ 0.05 mA to .999 mA* (Single Channel
Only)
1.00 mA to 9.99 mA
10.0 mA to 30.0 mA
Accuracy......................................................... 2 % of reading or ±50 µA (whichever is
greater)
Polarity Indicator ................................................. + or -
Rate .................................................................... 10 PPM to 99.9 PPM
100 PPM to 999 PPM
Accuracy......................................................... 0.5 % or 0.3 PPM (whichever is
greater)
Width................................................................... 0.02 mS to .999 mS
1.00 mS to 9.99 mS
10.0 mS to 99.99 mS
Accuracy......................................................... 0.5 % or ±14 µS (whichever is greater)
Resolution........................................................... ±1 LSD or ±4 µS (whichever is
greater)