Dc leakage current, Current drain test – Fluke Biomedical SigmaPace 1000 User Manual
Page 40
SigmaPace 1000
Operators Manual
1-26
DC Leakage Current
Test Type................................................................Quantitative
Measurement Range .............................................00.1 µA to 99.9 µA
Input Polarity..........................................................Positive and Negative
Resolution ..............................................................1 LSD
Display Format.......................................................3 digits plus decimal point
Test Selections
Static ...................................................................Continuous measurement
Pacemaker Power ...............................................OFF
Input Configurations ............................................Atrial+ and Atrial-
Ventricular+ and Ventricular-
Atrial+ and Ventricular- (third internal
500
Ω
test load)
Dynamic ..............................................................Gated measurement preceding the
delivered pacemaker pulse.
Pacemaker Power ...............................................ON
Input Configurations ............................................Atrial+ and Atrial-
Ventricular+ and Ventricular-
Atrial+ and Ventricular+ (third internal
500
Ω
test load)
Current Drain Test
Test Type................................................................Quantitative
Measurement Ranges
DC Current ..........................................................0.100 mA to .999 mA
1.00 mA to 9.99 mA
10.0 mA to 99.9 mA
Polarity ................................................................Positive or Negative
Polarity Indicator..................................................+ or - symbol preceding the numeric
value
Resolution ...........................................................±1 LSD
Display Format ....................................................3 digits plus decimal point
Accuracy..............................................................±5 % of Reading 10 µA
Input DC Voltage
Nominal ...............................................................±9 Volts
Range..................................................................5.0 Volts to 10.5 Volts
Input Protection ...................................................Short-circuit protection
Protection Type ...................................................Internal in-line fast-acting fuse rated at
½ Ampere.
Selectable Testloads ...........................................200
Ω
, 500
Ω
, and 1000
Ω