SigmaPace 1000

Operators Manual

1-26

DC Leakage Current

Test Type................................................................Quantitative
Measurement Range .............................................00.1 µA to 99.9 µA
Input Polarity..........................................................Positive and Negative
Resolution ..............................................................1 LSD
Display Format.......................................................3 digits plus decimal point
Test Selections

Static ...................................................................Continuous measurement
Pacemaker Power ...............................................OFF
Input Configurations ............................................Atrial+ and Atrial-

Ventricular+ and Ventricular-
Atrial+ and Ventricular- (third internal
500

Ω

test load)

Dynamic ..............................................................Gated measurement preceding the

delivered pacemaker pulse.

Pacemaker Power ...............................................ON
Input Configurations ............................................Atrial+ and Atrial-

Ventricular+ and Ventricular-
Atrial+ and Ventricular+ (third internal
500

Ω

test load)

Current Drain Test

Test Type................................................................Quantitative
Measurement Ranges

DC Current ..........................................................0.100 mA to .999 mA

1.00 mA to 9.99 mA
10.0 mA to 99.9 mA

Polarity ................................................................Positive or Negative
Polarity Indicator..................................................+ or - symbol preceding the numeric

value

Resolution ...........................................................±1 LSD
Display Format ....................................................3 digits plus decimal point
Accuracy..............................................................±5 % of Reading 10 µA

Input DC Voltage

Nominal ...............................................................±9 Volts
Range..................................................................5.0 Volts to 10.5 Volts
Input Protection ...................................................Short-circuit protection
Protection Type ...................................................Internal in-line fast-acting fuse rated at

½ Ampere.

Selectable Testloads ...........................................200

Ω

, 500

Ω

, and 1000

Ω