Long term test, Interactive pacer ecg simulation, Available testloads (500 – Fluke Biomedical SigmaPace 1000 User Manual
Page 41: Default)

Introduction and Specifications
Instrument
Specifications
1
1-27
Battery Test Fixture ............................................ Includes the 9 volt battery supply and
facilitates connection of Analyzer to
the recessed battery terminals within
the Medtronic 5388 and 5348
Temporary Pacemakers.
Long Term Test
Test Configuration................................................ Atrial Channel or Ventricular Channel
Pulse Count Range............................................... 999,999 (maximum)
Rate ........................................................................ 2 % to 20 % (10 % default)
Amplitude .............................................................. 2 % to 20 % (10 % default)
Test Time (max) .................................................... 999:59:59 (hhh:mm:ss)
Maximum error Count .......................................... 200
Test Termination................................................... Manual or when maximum error count
is sensed.
Testloads Selectable ............................................ 200 , 500 , or 1000
Interactive Pacer ECG Simulation
Simulates demand, continuous, non-capture, and non-function patient ECG
activity.
Additional user-selectable parameters
PR Interval .......................................................... 0.05 to 0.300 Seconds
Independent Capture / threshold ........................ 1 mA to 25 mA (1 mA steps) for each
pacemaker channel.
Available Testloads (500
Ω
Default)
Atrial Channel
Selections ........................................................... 200
Ω
, 500
Ω
, or 1000
Ω
Accuracy ............................................................. ±1.0 % of selection
Power Rating ...................................................... 2 Watts
Ventricular Channel
Selections ........................................................... 200
Ω
, 500
Ω
, or 1000
Ω
Accuracy ............................................................. ±1.0 % of selection
Power Rating ...................................................... 2 Watts
Tracking .............................................................. Atrial and ventricular channel settings
are identical.