Fluke Biomedical SigmaPace 1000 User Manual
Page 145

Remote Operation
Transvenous Pacemaker Remote Setup and Testing
5
5-15
RDA
RDV
The Analyzer transmits the following information for rate, amplitude, and
width from the atrial or the ventricular data “frozen” from the above HOLD
serial command. The data is returned in the following format:
AMP=XXX mA
RATE=XXX PPM
WIDTH=XXX mS
For each parameter, the Analyzer returns a three-digit numeric value (XXX)
with a floating point decimal to cover the multiple ranges. The unit of measure
remains the same for all transmitted readings (mA, PPM, and mS).
RDAH
Returns atrial amplitude and rate.
RDVH
Returns ventricular amplitude and rate.
RDAV
The Analyzer transmits the following information string for rate, amplitude,
and width from the atrial or the ventricular data “frozen” from the above
HOLD serial command.
The data is returned in the following format:
A Channel: AMP=XXX mA, RATE=XXX PPM, WIDTH=XXX mS
V Channel: AMP=XXX mA, RATE=XXX PPM, WIDTH=XXX mS
AV Interval: XXX mS
For each parameter, the Analyzer returns a three-digit numeric value (XXX)
with a floating point decimal to cover the multiple ranges. The unit of measure
remains the same for all transmitted readings (mA, PPM and mS)