Asynchronous mode test – Fluke Biomedical SigmaPace 1000 User Manual
Page 28

SigmaPace 1000
Operators Manual
1-14
Auxiliary Rate Range
Underdrive (minimum).........................................10 BPM
Overdrive (maximum)..........................................300 BPM
Active Outputs
Five lead ECG
Selected Ventricular Channel Test Load
High Level ECG Jack
This waveform is present on these three outputs.
Pacemaker Compatibility
Compatible Pulse Rates......................................30 to 200 PPM
Intended Type .....................................................External Transcutaneous VVI
(Ventricular Only Pace and Sense)
Asynchronous Mode Test
Test Type................................................................Qualitative
Physiological Simulation Selection
Selection..............................................................Normal Sinus Rhythm (NSR)
generates the complete P-QRS-T
complex.
Amplitude (V
peak
)..................................................1.0 mV Lead I
Rate (Interval)
Underdrive...........................................................NSR @ 85 % of measured pulse
interval / rate
Overdrive.............................................................NSR @ 115 % of measured pulse
interval / rate
Auxiliary Control ..................................................The under / overdrive NSR simulations
can be independently adjusted in 1
BPM increments.
Auxiliary Rate Range
Underdrive (minimum).........................................10 BPM
Overdrive (maximum)..........................................300 BPM
Active Outputs
Five lead ECG
Selected Ventricular Channel Test Load
This waveform is present on both outputs.
Pacemaker Compatibility
Compatible Pulse Rates......................................30 to 200 PPM
Intended Type
External Transcutaneous
VOO (Asynchronous Ventricular Only Pace)