Paced refractory period test (prp), Sensed refractory period test (srp) – Fluke Biomedical SigmaPace 1000 User Manual
Page 30

SigmaPace 1000
Operators Manual
1-16
Active Outputs
Selected Ventricular Channel Test load
Five lead ECG
Testload Amplitude Output
Ventricular Channel
Range..............................................................0.00 (OFF) to 10.0 mV
peak-to-peak
Accuracy..........................................................±5 % of setting
Resolution (Step Size).....................................0.50 mV
peak-to-peak
steps
Pacer Load Range ..........................................(30) 50
Ω
to 1550
Ω
1 %
Five Lead ECG Output
Range..................................................................0.00 (OFF) to 10.0 mV
peak-to-peak
Accuracy..............................................................±5 % of setting
Resolution (Step Size).........................................0.50 mV steps
Calibration Reference..........................................Lead I (RA to LA)
Paced Refractory Period Test (PRP)
Range......................................................................20 mS to 500 mS
Accuracy..............................................................5 % of reading or 1 mS whichever is
greater
Physiological Simulation
Selection..............................................................Triangle (TRI) Wave
Pulse Width .........................................................40 mS
Rate.....................................................................Single pulse: Interactive with applied
pacemaker pulse activity in the
demand mode of operation.
Amplitude ...............................................................1.0 mV
peak
Lead I
Active Outputs .....................................................Five Lead ECG and selected
ventricular channel test load This
waveform is present on both outputs.
Pacemaker Compatibility
Compatible Pulse Rates......................................30 to 200 BPM
Intended Pacemaker Type ..................................VVI (Ventricular Pace and Sense
Only)
Sensed Refractory Period Test (SRP)
Range......................................................................15 mS to 500 mS
Accuracy..............................................................±5 % of reading or ±1 mS, whichever
is greater
Physiological Simulation
Selection..............................................................Triangle (TRI) Wave