Fluke Biomedical SigmaPace 1000 User Manual
Page 39

Introduction and Specifications
Instrument
Specifications
1
1-25
Resolution ............................................................. ±1 LSD
Display Format...................................................... 3 digits
Physiological Simulation
Selection (Default) .............................................. Single Channel Simulation (V)
Square Wave
Ventricular Channel ............................................ Simulated R-wave
Width................................................................... 1.0 mS
Amplitude............................................................ 20 mV
peak
Active Outputs .................................................... Ventricular Channel
(4 mm Banana Jacks) Only
Available Test Load(s) ........................................ 500
Ω
Only
Additional Waveform Selections
Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO) (Fixed Width2 mS rise time / 13 mS fall time)
Amplitude
Pacer Load Selection.......................................... 500
Ω
Range ............................................................. 0.05 mV
peak
(50 V
peak
) to
50.0 mV
peak
Accuracy......................................................... ±5 % of setting
Resolution (Step Size).................................... 0.05 mV
peak
steps from 0.05 mV
peak
to
0.95 mV
peak
0.50 mV
peak
steps from 1.0 mV
peak
to
50.0 mV
peak
Default Setting ................................................ 20.0 mV
peak
Width
Range ................................................................. 0.15 mS to 300 mS
Accuracy ............................................................. 5 % of setting
Selection Count .................................................. 58
Resolution (Step Size)........................................ 0.05 mS steps from 0.15 mS to
0.95 mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS
25 mS steps from 100 mS to 300 mS
Default Setting (mS) ........................................... 30 mS
Intended Pacemaker Types................................ VVI (Ventricular Pace and Sense
Only)
Compatible Pacemaker Rates ............................ 20 to 200 PPM