Remote Operation
Transcutaneous Pacemaker Remote Setup and Testing
5
5-7
SETMAKE
SETMAKE=nnnn
The specific protocol required for each manufacturer (or make) is selected
using this command. Each protocol contains a set of measurement algorithms
for rate, width, amplitude, and refractory. SETMAKE= followed by the four-
digit alphanumeric entry assigned to each manufacturer completes this serial
command.
The following table shows the SETMAKE protocol for the indicated
manufacturers (Example: SETMAKE = ZM07).
Table 5-1. SETMAKE Protocols by Manufacturer
Manufacturer (Make)
Selection
Default/Physio-Control/Cardiotronics PC01
Hewlett Packard Codemaster
HP02
Marquette Medical
MQ03
Medical Data Electronics
MD04
Medical Research Laboratories
MR05
Zoll Medical PD1200 – 2000
ZP06
Zoll Medical M-Series
ZM07
RDMAKE
This command returns the NAME and Reference # for the specified make.
(Example: ZM-07 = Zoll Medical M-Series pacemaker)
SETLOAD
SETLOAD=nnnn
This command is used to select the desired test load. The above “nnnn”
represents the actual test load resistance value (in ohms) and is based upon the
available factory preset selections.