Fluke Biomedical SigmaPace 1000 User Manual
Page 139

Remote Operation
Transcutaneous Pacemaker Remote Setup and Testing
5
5-9
RDAMP
The Analyzer transmits the following information string for amplitude from
the data “frozen” with the above HOLD serial command. The data is returned
in the following format: AMP=XXXmA
RDRATE
The Analyzer transmits the following information string for rate from the data
“frozen” with the above HOLD serial command. The data is returned in the
following format: RATE=XXXPPM
RDWIDTH
The Analyzer transmits the following information string for width from the
data “frozen” with the above HOLD serial command. The data is returned in
the following format: WIDTH=XXXmS
RDENERGY
Even though the pacemaker pulse energy reading is not displayed when the
RDMETER serial command is utilized, it is computed. The Analyzer transmits
information from the data “frozen” with the HOLD serial command.
One of the following formats is used, depending on the energy level of the
transcutaneous pacemaker pulse.
ENERGY= XXX?J
ENERGY=XXXmJ
ENERGY=X.XXJ
RESUME
Releases the HOLD serial command and proceeds with making measurements.
SETTEST=DEMAND
The Analyzer enters the qualitative demand test. The applied pacemaker pulse
rate is measured; if it is ZERO, the Analyzer transmits RATE=000 and will not