2.1.5 Unit Protection
Input Over current:
In-line fast acting fuses. Check fuse rating in Service and Maintenance section.
Ratings will depend on AC input configuration settings.
Circuit breaker for LV supply.
Input Over voltage:
Automatic shutdown.
Input Over voltage
Transients:
Surge protection to withstand EN50082-1 (IEC 801-4, 5) levels.
Output Over current:
Adjustable level constant current mode with programmable set point.
Output Short Circuit:
Peak and RMS current limit.
Over temperature:
Automatic shutdown.
2.2 Mechanical
Parameter
Specification
Dimensions:
(for each BPS 30 / BPS45
chassis)
Height: 50.0”
1270
mm
Width: 28.75”
731
mm
Depth: 34.5”
876
mm
Unit Weight:
(for each BPS30/BPS45
chassis)
Net: 1150 lbs / 522 Kg approximately
Shipping: 1231 lbs / 560 Kg
approximately
Dimensions:
(for each BPS75 / BPS90
chassis)
Height: 74.5”
1892.3 mm
Width: 30.3”
769.6 mm
Depth: 38.3”
972.8 mm
Unit Weight:
(for each BPS75 / BPS90
chassis)
Net: 2150 lbs / 975 Kg approximately
2475 lbs / 1123 Kg approximately
Shipping: 2450 lbs / 1111 Kg
approximately
2775 lbs / 1258 Kg approximately
Material:
Steel or aluminum chassis, panels and covers.
Finish:
Light textured painted external surfaces.
Panels semi-gloss polyurethane color no. 26440 (medium gray)
Cooling:
Fan cooled with air intake on the front and exhaust to the rear.
Fans:
14 x 225CFM.
Air displacement
50 Cu Ft/sec. Max.
Internal Construction:
Modular sub assemblies.
Rear Panel
Connections:
(See section 3and 4 for description of connections)
Cable entry and strain relieve for AC input wiring
Cable entry and strain relieve for AC output wiring
External sense terminal block (Remote voltage sense)
System interface (2x)
RS232, USB, GPIB, LAN (option)
Trigger In BNC
Trigger Out BNC
Function Strobe BNC
Output Status