7.2 Front Panel Calibration Screens

The calibration screens for output or measurement calibration can be selected from the MENU 3 screen.
(Press MENU button several times to toggle to MENU 3 screen.)

To select the OUTPUT CALIBRATION screen press the

 or  key several times to highlight OUTPUT

CAL. Then press the ENTER key. This will bring up the PASSWORD screen. To prevent unauthorized
access to calibration data, a password must be entered to access any calibration screen. The calibration
password is always "5000" and may be entered using the numeric keypad. Once entered, the calibration
screens remain accessible until the BPS unit is powered down.

Type 5000 and press the ENTER key to show the OUTPUT CALIBRATION screen.

On BPS systems with three-phase output capability, use the PHASE key on the front panel to select the
phase to be calibrated.

To select the MEASUREMENT CALIBRATION screen, follow the same steps as outlined above but
select the MEASUREMENT CAL entry instead of OUTPUT CAL. If another CALIBRATION screen
has been accessed since power-up, no password is needed. Otherwise, enter the same password as
indicated above.