5.5.4 Pulse Transients

Pulse transients let you program the output to a specified value for a predetermined amount of time. At the
end of the Pulse transient, the output voltage returns to its previous value. Parameters required to set up a
Pulse transient include the pulse count, pulse period, and pulse duty cycle. An example of a Pulse transient
is shown in Figure 5-28. In this case, the count is 4, the pulse period is 16.6 ms or 60 Hz and the duty
cycle is 33%.

Figure 5-28: Pulse Transients

Note that Pulse transients can only be programmed over the bus, not the front panel. Refer to the SCPI
Programming Manual for more information about programming Pulse transients and triggers.