AMETEK Programmable Power

BPS Series

122

Note: The Agilent 3458A Digital Multimeter or equivalent must be used for the following calibration.

The shunt must be connected to the power source as shown in Figure 7-2. If the Current Measurement
can’t be successfully performed, adjust the Current Measurement Pot on the System Interface board. This
adjustment is described in the Non-routine Calibration section of this manual. If the DC current
measurement displays more than 70 counts on the display, perform the non-routine current monitor
adjustment.