beautypg.com

AMETEK BPS Series User Manual

Page 128

background image

User Manual

AMETEK Programmable Power

BPS Series

128

7.4.2 Three Phase Mode

As indicated earlier, for 3-Phase power system (BPS30-3, BPS45-3, BPS75-3, BPS90-3, BPS150-3 and
BPS180-3), repeat the preceding steps for the Phase B and C outputs. The order in which the outputs for
each phase are calibrated is not important.

Press the PHASE key to select each output to be calibrated. Monitor the output of the respective phase by
moving the HI input of the Digital Multimeter. The LO input should remain connected to the common LO
of the sense connector.

7.4.3 Phase Angle Calibration

Output phase angle calibration is required only on BPS Series models capable of three-phase mode
operation. A phase meter is required to perform this calibration.

This calibration can be done in either high or low voltage range as long as the maximum input voltage of
the phase meter input is not exceeded. Select AC mode, thee-phase mode, 120Vac. 60 Hz. Program all
three-phase outputs to the same voltage.

For best results, the output of all amplifiers should be loaded to at least 80% of full scale voltage current
on all phases at the same time. This requires a three-phase load. If no such load is available, perform this
calibration with no load.

Proceed as follows:

1. Program AC mode, 3 phase, low voltage range, 120Vac, 50 or 60 Hz. Close output relay.

2. Select the OUTPUT CAL screen.

3. Connect phase meter between phase A and B outputs.

4. Use the PHASE key to select phase B in the upper right corner of the CAL screen.

5. Close output relay and measure the phase angle between phase A and Phase B.

6. Adjust PHASE OFST cal coefficient up or down and press ENTER key until phase B offset is

240° ± 0.5° or better.

7. Connect phase meter between phase A and C outputs.

8. Use PHASE key to select phase C in the upper right corner of the CAL screen.

9. Adjust PHASE OFST cal coefficient up or down and press ENTER key until phase C offset is

120° ± 0.5° or better.

7.4.4 Output Calibration Summary

The following Table is a summary of the preceding calibration steps. Program the following values in the
table and make the adjustments in the OUTPUT CALIBRATION screen. Select the phase to be calibrated
by pressing the PHASE key.

CALIBRATION

PROGRAM VALUES

CALIBRATION VALUE

ADJUST TO

300 VAC range DC Zero

300 VAC range, 0.0 V,
60 Hz

VOLT ZERO

0 ± 5 mV DC

300 VAC range High Freq DC
Zero, -HF models only.

300 VAC range, 0.0 V,
820 Hz

VOLT ZERO

0 ± 5 mV DC

300 VAC range Volt FS

240.0 V, 60 Hz

VOLT FS

240

 0.05 VAC

Phase Offset B, C

150 VAC range, 120V,
60 Hz

PHASE OFST

± 0.5

°

Table 7-3: Output Calibration Table – BPS Series