Concurrent balance values – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 88
Product Service 1-800-522-7658
88
Concurrent Balance Values
Axiom Inflow and Inlet Valve flow numbers are established by a direct measurement of
air flow. Traditional exhaust system measurements are made using either an
anemometer or Pitot traverse methodology. Because of these different methodologies,
the total volume of air flowing through the exhaust system will vary.
Table 5-6
Concurrent Balance Volume Specifications
Model
30448
30441
30468
30461
DIM-measured Inflow (CFM)
283
354
423
529
Nominal Inflow through Valve (CFM)
40
40
40
40
Total DIM-measured (CFM)
323
394
473
569
Traverse-measured Exhaust (CFM)
387
480
556
684