Quickstart, Main menu 1. mylogic, Set t ings 3. service – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 12: Service
Product Service 1-800-522-7658
12
Quickstart
How do I see the Current Settings of the Axiom?
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted (it will
turn yellow). Press
OK/Mute
to enter the Services menu
screen.
3.
Press the
▼
button until the
Current Settings
option is
highlighted (it will turn white). Press
OK/Mute
to see the
configuration.
4.
The first screen shows the current settings of the supply blower
system.
5.
By pressing the
▲
or
▼
button, you can see the current
settings of the exhaust blower system. If you are recertifying the
unit, you may want to record the data on this screen. The
firmware revision is shown in the bottom right corner as a -, or
as a letter.
MAIN MENU
1. MyLogic
TM
2. Set t ings
3. Service
or
t o choose,
OK t o select , MENU t o escape
or
t o choose,
OK t o select , MENU t o ret urn
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
Service
Current Set t ings
Supply Syst em
or
t o change, OK t o ret urn
The ECM sett ing is: 65
The current RPM is: 0
The average RPM is: 0
The new filter RPM is: 1450
C1, wit h afs, with UV
10484 -
Current Set t ings
Exhaust Syst em
or
t o change, OK t o ret urn
The ECM sett ing is: 65
The current RPM is: 0
The average RPM is: 0
The new filter RPM is: 1450
C1, wit h afs, with UV
10484 -