Quickstart
How do I run a diagnostic on the keypad, relays and contacts?
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted (it will
turn yellow).
3.
Press
OK/Mute
to enter the Services menu screen.
4.
Press the
▼
button until the
Diagnostic
option is highlighted (it
will turn white).
5.
Press
OK/Mute
to enter the Diagnostic submenu.
6.
Press the
▼
button until the Diagnostic
Keypad-Relays
option is highlighted
(it will turn white). Press
OK/Mute
.
7.
Press any button on the keypad (except
Menu
), and note the
display.