Quickstart
How do I recalibrate the
blower speed
filter life gauge
airflow sensor reading
of the Axiom? - Continued
10.
If this is an initial certification or a recertification after supply
HEPA filter replacement, select
Reset Filter Gauge to 100%
. If
this is an annual recertification, select
Leave Gauge unchanged
.
If you want to set the filter gauge at a preset value, select
Set
Gauge to new setting
. Then press
OK/Mute
.
11.
The next certification screen will display the exhaust blower
motor’s PWM setting. Press the
▲
or
▼
button to increase or
decrease the blower speed as needed. When the speed is set
properly, press the
OK/Mute
button to lock the new blower
speed setting, and continue to the filter life screen.
12.
If this is an initial certification or a recertification after exhaust
HEPA filter replacement, select
Reset Filter Gauge to 100%
. If
this is an annual recertification, select
Leave Gauge unchanged
.
If you want to set the filter gauge at a preset value, select
Set
Gauge to new setting
. Then press
OK/Mute
.
13.
If the unit is has no airflow sensor, after the filter gauge option
is selected and you press
OK/Mute
, the Axiom will power
down, and then restart with the new settings. If the Unit has an
airflow sensor, you will then be asked to calibrate the inflow
sensor to match the value you obtained when you measured the
inflow. After the inflow is calibrated, and you press
OK/Mute
,
the Axiom will power down, and then restart with the new
settings.
The Inflow displayed
value is: 102 FPM
Calibrat ion
Inf low Display
or
t o change display value,
UV Light t o reset t o 105 FPM,
OK t o select , MENU t o f inish
Calibration
Supply Filter Gauge
or
t o change f ilt er gauge,
OK t o select , MENU t o escape
Reset Filt er Gauge t o 100%
Leave Gauge unchanged
Set Gauge t o new set t ing
The ECM set t ing is: 80
The current RPM is: 1050
The average RPM is: 1020
Calibration
Exhaust Fan Speed
or
t o change mot or speed,
OK t o select , MENU t o escape
Calibration
Exhaust Filter Gauge
or
t o change f ilt er gauge,
OK t o select , MENU t o escape
Reset Filt er Gauge t o 100%
Leave Gauge unchanged
Set Gauge t o new set t ing