Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 54
Product Service 1-800-522-7658
54
Disconnecting the Axiom from an Exhaust System Connection to Operate in
Reciculation
a)
Mechanical configuration
1.
Unplug the Axiom. Using an appropriate ladder or platform, remove the
exhaust cover panel on the top left side of the exhaust cover, and install it
on the studs on the right side. Hand tighten the wing nuts to secure it, as
shown in figure 3-1a.
2.
Disconnect the Axiom’s connection collar from the exhaust system, and cap
the building’s exhaust connection in a way that meets all appropriate codes
and regulations. Leave the Axiom’s exhaust collar open to allow for
unrestricted airflow out of the unit.
Exhaust Cover Panel
Figure 3-1a
This manual is related to the following products: