Note: Any other key sequence will return you to the Attention screen.
9.
The first screen allows you to select whether the
Axiom is connected to an exhaust system or not.
IT IS CRITICAL THAT THE CABINET
MATCHES THE OPTION THAT IS SELECTED.
Press
OK
to continue.
10.
If connected to an exhaust system, the second
screen allows you to set how long the blowers in
the Axiom will continue to operate in the event of
an exhaust system failure. The interval can be
programmed from 0-300 seconds.
Note: Consult with your facility’s safety officer or Labconco to help establish how long
the Axiom should continue to operate after an exhaust system failure.
11.
The third screen allows you to set the sash height
at either 8 or 10 inches. IT IS CRITICAL THAT
THE YOU LEAVE THE SASH HEIGHT AS IT
WAS SET AT THE FACTORY, UNLESS YOU
WISH TO RECERTIFY THE CABINET AT ITS
NEW SETTING.
12.
The fourth screen allows you to select whether the
Axiom has a UV light or not. IT IS CRITICAL
THAT THE YOU SELECT
A UV LIGHT
. Press
OK
to return to the first Configuration screen.
The unit is now properly configured for operation with the UV light.
Refer to the user’s manual for further information regarding the proper
operation and maintenance of the UV light.