Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 52

Product Service 1-800-522-7658
52
3.
The screen will now prompt you for the password;
it is:
Light
UV Light
Timer
Timer
OK
Note: Any other key sequence will return you to the
Attention screen.
4.
The first screen allows you to select whether the
Axiom is connected to an exhaust system or not.
IT IS CRITICAL THAT THE
THIS CABINET
HAS AN EXHAUST CONNECTION
OPTION IS
SELECTED. Press
OK
to continue.
5.
The second screen allows you to set how long the
blowers in the Axiom will continue to operate in the
event of an exhaust system failure. The interval
can be programmed from 0-300 seconds.
Note: Consult with your facility’s safety officer or Labconco to help establish how long
the Axiom should continue to operate after an exhaust system failure.
6.
The third screen allows you to set the sash height at
either 8 or 10 inches. IT IS CRITICAL THAT THE
YOU LEAVE THE SASH HEIGHT AS IT WAS SET
AT THE FACTORY, UNLESS YOU WISH TO
RECERTIFY THE CABINET AT ITS NEW
SETTING.