Hepa leak test information, Supply hepa filter leak test – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 76

Product Service 1-800-522-7658
76
HEPA Leak Test Information
All Axiom models were tested to the HEPA Filter Leak Test as described in ANSI/NSF
Standard 49:2012.
Note: ensure all internal surfaces of the Axiom have been appropriately
decontaminated before proceeding.
Supply HEPA Filter Leak Test
All Axiom models were tested to the HEPA Filter Leak Test as described in ANSI/NSF
Standard 49:2012.
Note: ensure all internal surfaces of the Axiom have been appropriately
decontaminated before proceeding.
Note: For all Axiom models, the work surfaces may be removed if desired, and
the aerosol generator discharge should be placed to the left of (but NOT in)
the center exhaust duct nearby to inject the aerosol as shown in Figure 5-1.
Figure 5-1
Place aerosol generator discharge into
the left recirculation duct
Center exhaust duct