Preparing the axiom for gaseous sterilization – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 112

Product Service 1-800-522-7658
112
Preparing the Axiom for Gaseous Sterilization
Note: This section only reviews the steps required for preparing the Axiom for
Gas Sterilization. Thoroughly understand the sterilization procedures and
protocols supplied by the manufacturer of the sterilizing system before
attempting this operation.
1.
Unplug the Axiom from electrical power.
2.
Thoroughly surface decontaminate the working area of the cabinet.
In order to get a complete seal on the front of the cabinet, do the following:
3.
Remove the work surfaces after it has been surface decontaminated.
4.
Remove the front grille.
5.
Place the work surfaces and the grille into the work area for
decontamination.
6.
Remove the front dress panel.
7.
Remove the sash stop bracket, located near the top of the blower plenum
cover.
8.
Lift the sash up until the bottom edge of the sash clears the bottom edge of
the Supply HEPA Filter shelf.
9.
Using flexible plastic sheet 5-8 mil thick and duct tape seal the perimeter of
the front opening.
10.
Using flexible plastic sheet and duct tape seal the Exhaust filter opening.
11.
The Axiom is now prepared for gaseous decontamination.