Inflow velocity calculation - primary method – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 84
Product Service 1-800-522-7658
84
Inflow Velocity Calculation - Primary Method
Note: The dress panel MUST be installed on the cabinet to obtain an accurate inflow
value.
The average inflow velocity for Axiom models is 105 ±5 FPM. It should be determined
by converting the inflow volume to the average inflow velocity. The inflow volume is
measured directly by sealing a direct inflow measuring flow meter to the face of the
unit. When corrected for local conditions, dividing the inflow volume by the opening area
will yield the average inflow velocity.
Table 5-4
Inflow Test Specifications – 8 inch Sash Opening
Width (feet)
4
6
Nominal Average Inflow
(FPM)
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
2.69
4.03
Nominal Avg. Exhaust Vol.
(CFM)
283
423
Avg. Exhaust Vol. Range
(CFM)
269-296
403-443
Inflow Test Specifications – 10 inch Sash Opening
Width (feet)
4
6
Nominal Average Inflow
(FPM)
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
3.37
5.03
Nominal Avg. Exhaust Vol.
(CFM)
354
529
Avg. Exhaust Vol. Range
(CFM)
337-371
503-553
Acceptance
Average inflow shall be meet the specified value.