Quickstart
How do I change the Configuration of the Axiom? – continued
9.
The first screen allows you to decide if the cabinet is connected to
an exhaust system or not. When the unit is configured for no
exhaust connection, the Axiom will ignore the inlet relief valve
sensor’s status. Selecting connected to an exhaust system
activates the sensor; if the valve is open, the unit operates
normally, if closed, an exhaust alarm is activated. Use the
▲
or
▼
buttons to select the exhaust configuration, and then press
OK/Mute.
10.
If the unit is configured to be connected to an exhaust system,
then the next screen allows you to program the amount of time
interval of Active Protection. Active Protection will allow the Axiom
blowers to continue to operate after the detection of an exhaust
system failure. This will maintain a safe inflow to allow the user a
programmed time interval to secure any hazardous materials in
the cabinet before the BSC shuts off.
11.
This screen configures the sash height. After your selection is
chosen, and you press
OK/Mute
.
12.
This screen configures the unit for a UV light or not. After your
selection is chosen, and you press
OK/Mute
, the unit will shut off
and then reset in the new configuration.
13.
If the unit is configured to NOT have a UV light, and the user
presses the UV Light button, they will get this message.
This cabinet is not configured
for a UV light
If you want to install a UV
light, contact Labconco Product
Service at 800- 821- 5525
At t ent ion
MENU t o ret urn
Conf igurat ion
Type C Conf igurat ion
This cabinet has:
an
or
t o choose,
OK t o select , MENU t o escape
no exhaust connect ion
exhaust connect ion
Conf igurat ion
Type C Conf igurat ion
If t here is an exhaust syst em
f ailure, Act ive prot ect ion should
last :
or
t o choose,
OK t o select , MENU t o escape
0 Seconds
The sash height is:
or
t o choose,
OK t o select , MENU t o escape
8 inches
10 inches
NOTE: If you change t he sash
height , you must recert if y
airf lows.
Sash Height Settings
This Logic has:
or
t o choose,
OK t o select , MENU t o escape
a UV light
NO UV light
UV Settings