11.2.2 Continuous meter log

• Select “Start meter log” to log the currently en-

abled parameters at the set logging interval on
the meter.

• To set the logging interval, highlight “Start meter

log” and press . The log interval time
can set from 1 second to 3 hours. Press
and use the arrow keys and keypad
to enter the desired log interval. Press
to confirm.

• Press . Enter the desired time
and press to confirm.

• Enable feature to overwrite

the oldest records in the current log. (Note: If the