beautypg.com

Hanna Instruments HI 9829 User Manual

Page 39

background image

39

• After the calibration point is confirmed, to avoid cross-contamination immerse

the sensors in the next calibration buffer rinse solution and stir gently.

• Press to select the next buffer (if necessary), and repeat the cali-

bration procedure outlined above with the second and third buffers.

Note

Note

Note

Note

Note

The calibration procedure can be terminated after a single or 2 point
calibration by pressing . The message “Storing” followed by
“Calibration completed” will be displayed.

• Press to return to the Calibration menu.
• Press to return to the measurement screen.
Note: If a single point calibration is conducted after restoring the factory calibra-

tion, it will offset the entire factory probe values by the same offset required to
calibrate the sensor to the single buffer value.

Custom buffer calibration
The HI 9829 permits a single custom buffer to be used for pH calibration. This
can be used along with standard buffers as part of a 2 or 3 point calibration or
as a single point.
• To select this option first press and then while the

meter is waiting for stable reading.

• A text box window will appear. Use the keypad to enter the value of the buffer

at the current temperature. The valid range for custom a buffer is from 0.00
to 14.00 pH.