If the % D.O. saturation range is calibrated, the D.O. concentration range will
also be calibrated, and vice versa.
Dissolved oxygen concentration values are based
on % D.O. saturation, temperature, salinity and
atmospheric pressure. A standard solution or a
reference D.O. meter may be used to compare
readings during calibration.
The calibration of the D.O. concentration range
can only be performed at a single custom point (4 to 50 mg/L). It is recommended
to calibrate the D.O. sensor close to the values that will be measured.
Choose “DO calibration” from the “Calibration” menu, select the D.O. calibration
type using the arrow keys and press