beautypg.com

Hanna Instruments HI 9829 User Manual

Page 52

background image

52

7.9 ATMOSPHERIC PRESSURE CALIBRATION
Place HI 9829 in a wind-free area and choose “Custom pressure” to perform a

user calibration or “Restore factory calib”.

Note

Note

Note

Note

Note

“Custom pressure” procedure requires a reference barometer.

Select the “Atm. pressure” from the “Calibration” menu.
• Select the “Custom pressure” option.
• Using the keypad, insert the numeric value that agrees with the reference meter

and then press to confirm.

• The stability counter will count down and the message “Ready” and “Confirm”

will be displayed. Press to store the calibration point.

• After confirmation, the following messages are displayed: “Storing” and “Cali-

bration completed”.

• Press to return to the measurement screen.
• Press to return to the “Calibration” menu.
• To restore the factory calibration, select “Restore factory calib.” in the “Pressure

calibration” menu and press .