7.9 ATMOSPHERIC PRESSURE CALIBRATION
Place HI 9829 in a wind-free area and choose “Custom pressure” to perform a
user calibration or “Restore factory calib”.
Note
Note
Note
Note
Note
“Custom pressure” procedure requires a reference barometer.
Select the “Atm. pressure” from the “Calibration” menu.
• Select the “Custom pressure” option.
• Using the keypad, insert the numeric value that agrees with the reference meter
and then press
• The stability counter will count down and the message “Ready” and “Confirm”
will be displayed. Press
• After confirmation, the following messages are displayed: “Storing” and “Cali-
bration completed”.
• Press
• Press
• To restore the factory calibration, select “Restore factory calib.” in the “Pressure
calibration” menu and press
• After confirmation, the following messages are displayed “Storing” and “Cali-
bration completed”.
• Press
• Press
• To restore the factory calibration, select the corresponding option in the
“Temperature calib.” menu and then press