Notes A “C” label near the calibration point indicates a custom point, while

an “H” indicates a HANNA standard value.
When the % D.O. range is calibrated, also the D.O. concentration range
is calibrated, and vice versa.
If no D.O. calibration has been performed or if calibration was cleared
using the “Restore factory calib.” option the offset and slope values
are set to default, and the message ”Factory calibration” is displayed.
Press to return to the previous screen.

Conductivity
• From the “GLP” menu select the “Conductivity”

option.

• Data regarding the last conductivity calibration

will be displayed: calibration point, cell constant
value, calibration type (conductivity, absolute
conductivity or salinity), time and date of the cali-
bration.

• Use the arrow keys to scroll through the stored data for the last 5 calibrations.

Notes

Notes

Notes

Notes

Notes A “C” letter near the conductivity calibration indicates a custom point,

while an “H” indicates a HANNA standard value.
If no conductivity calibration has been performed or if calibration was
cleared using the “Restore factory calib.” option the offset and slope
values are set to default, and the message ”Factory calibration” is dis-
played. Press to return to the previous screen.

Turbidity
• From the “GLP” menu select the “Turbidity” option.
• Data regarding the last turbidity calibration

will be displayed: standards used, time and
date of the calibration.

• Use the arrow keys to scroll through the

stored data for the last 5 calibrations.

• Press to return to the “GLP” menu.