Hanna Instruments HI 9829 User Manual
Page 46

46
found in the bottom of the conductivity sensor. Clean using the small brush from
the probe maintenance kit. Flush with water. A mild detergent may be used to
remove oily coatings. Always flush with clean water after cleaning.
Note
Note
Note
Note
Note For a correct conductivity calibration, the probe shield or the calibration
beaker must be used.
The conductivity calibration menu includes 3
different types of calibration: Conductivity, Absolute
conductivity and Salinity.
The “Conductivity” option allows a single point
calibration with a standard solution selectable by
the user. This calibration is temperature
compensated.
The “Absolute conductivity” option allows a single point calibration with a
conductivity solution of known non-temperature compensated value at the current
temperature.
The “Salinity” option allows calibration with a standard salinity solution.
The 3 calibrations are related, so that each one will calibrate all 3 measurements.
Note
Note
Note
Note
Note
To improve accuracy, choose a calibration standard near the sample
conductivity.
Choose “Conductivity calibration” from the “Calibration” menu, select the
calibration type using the arrow keys and press
cessories” for choosing the proper HANNA standard solution).
• Pour additional standard into a second beaker to be used to rinse the sensor.
• Immerse the sensor into the rinse standard by raising and lowering the beaker
a few times to ensure that the EC sensor channels are filled with fresh stan-
dard.
• Place the calibration beaker over the EC sensor and dislodge any trapped
bubbles. Screw the beaker into place. Wait for the reading to stabilize.